Navigator-triggered and breath-hold 3D MRCP using compressed sensing: image quality and method selection factor
Daisuke Morimoto1, Tomoko Hyodo2, Ken Kamata3
1Radiology Center, Kindai University Hospital, 377-2 Ohno-Higashi, Osaka-Sayama, Osaka, Japan. morimoto-d@med.kindai.ac.jp.
Purpose:
To examine whether MRCP using a combination of compressed sensing and sensitivity encoding with navigator-triggered and breath-hold techniques (NT C-SENSE and BH C-SENSE, respectively) have comparable image quality to that of navigator-triggered MRCP using only sensitivity encoding (NT SENSE) at 1.5-T.
Methods:
Fifty-one participants were enrolled in this prospective study between July and October 2018 and underwent the three 3D MRCP sequences each. The acquisition time and relative duct-to-periductal contrast ratios (RC values) of each bile duct segment were obtained. Visualization of the bile and main pancreatic ducts, background suppression, artifacts, and overall image quality were scored on 5-point scales. Mean and median differences in RC values and qualitative scores of NT C-SENSE and BH C-SENSE relative to NT SENSE were calculated with 95% confidence intervals (CIs).
Results:
Acquisition time of NT SENSE, NT C-SENSE, and BH C-SENSE were 348, 143 (mean for both), and 18 s (for all participants), respectively. The RC value of each bile duct segment was inferior, but the lower limits of the 95% CIs of the mean differences were ≥ - 0.10, for both NT C-SENSE and BH C-SENSE. The visualization score of the intrahepatic duct in BH C-SENSE was inferior to that in NT SENSE (lower 95% CI limit, - 1.5). In both NT C-SENSE and BH C-SENSE, the 95% CIs of the median differences in the other qualitative scores were from - 1.0 to 0.0.
Conclusion:
NT C-SENSE and BH C-SENSE have comparable image quality to NT SENSE at 1.5-T.