Calcium Phosphate Bone Void Filler Increases Threaded Suture Anchor Pullout Strength: A Biomechanical Study.
Miguel A Diaz1, Eric A Branch2, Luis A Paredes1
1Foundation for Orthopaedic Research & Education, Tampa, Florida, U.S.A.
Summary
Augmenting threaded suture anchors with calcium phosphate bone substitute material significantly improves pull-out strength in osteoporotic bone models. This enhancement offers greater resistance to failure, addressing concerns about anchor security in weakened bone.
Area of Science:
- Orthopedic biomechanics
- Biomaterials science
- Surgical implant technology
Background:
- Suture anchor pull-out from osteoporotic bone is a significant clinical concern.
- Investigating methods to enhance the fixation strength of orthopedic implants is crucial for improving patient outcomes.
- Calcium phosphate bone substitute materials offer potential for bone void filling and augmentation.
Purpose of the Study:
- To compare the cyclic loading response and ultimate pull-out strength of threaded suture anchors.
- To evaluate the efficacy of calcium phosphate bone substitute material (CP-BSM) augmentation.
- To assess augmentation in both a polyurethane foam block model mimicking osteoporotic bone and an in vitro cadaveric humerus model.
Main Methods:
- A controlled biomechanical study involving two models: polyurethane foam blocks and proximal humerus cadaveric specimens.
- Foam blocks and cadaveric bone were either augmented with injectable CP-BSM or left untreated.
- Specimens were instrumented with threaded suture anchors and subjected to cyclic axial loading (40 cycles) followed by load-to-failure testing.
Main Results:
- Augmentation with CP-BSM significantly increased average pull-out strength in the osteoporotic foam block model (332.68 N vs. 144.38 N, P = .005).
- Similarly, CP-BSM augmentation significantly enhanced average pull-out strength in the cadaveric humerus model (274.07 N vs. 138.53 N, P = .029).
- Failure modes were analyzed, with augmentation demonstrating improved load-to-failure resistance.
Conclusions:
- Augmentation of osteoporotic bone models with calcium phosphate bone substitute material significantly increases the pull-out strength of threaded suture anchors.
- CP-BSM augmentation provides enhanced load-to-failure resistance, addressing clinical concerns regarding suture anchor fixation in osteoporotic bone.