Related Experiment Video
Updated: Dec 29, 2025

Magnetic Resonance-Guided Stereotaxy for Infusions to the Pig Brain
Published on: March 31, 2023
Relationship between pin type and depth of skull penetration during frame placement for Gamma Knife radiosurgery
Ronald E Warnick1, Eunsun Yook2
1Mayfield Clinic and the Gamma Knife Center, Jewish Hospital-Mercy Health, 3825 Edwards Road, Suite 300, Cincinnati, OH, USA.
Purpose:
Our retrospective study addresses the largely unknown effect of fixation pin type on depth of skull penetration during Gamma Knife frame placement.
Methods:
Absolute and relative depths were compared for 404 pins of aluminum and titanium in 101 patients who underwent Gamma Knife frame placement at 0.4 Nm (3.5 lb-in) torque.
Results:
Effect of pin type was significant: penetration was less in aluminum than titanium pins, for absolute (0.27 mm vs. 0.83 mm) and relative (5.16% vs. 15.53%) depths (p < 0.001 each). Fifty percent of aluminum pins did not measurably penetrate the outer table of the skull whereas 95% of titanium pins penetrated into cancellous bone.
Conclusions:
Although these pin penetration differences did not have a clear clinical correlate, we now recommend titanium pins for their consistent, deeper penetration, which may result in greater frame stability favorable for most clinical applications. Future studies may define optimal torque for aluminum pins.