Automatic detection of complete and measurable cardiac cycles in antenatal pulsed-wave Doppler signals
Eleonora Sulas1, Monica Urru2, Roberto Tumbarello2
1Department of Electrical and Electronic Engineering, University of Cagliari, Italy.
Background And Objective:
Pulsed-wave Doppler (PWD) echocardiography is the primary tool for antenatal cardiological diagnosis. Based on it, different measurements and validated reference parameters can be extracted. The automatic detection of complete and measurable cardiac cycles would represent a useful tool for the quality assessment of the PWD trace and the automated analysis of long traces.
Methods:
This work proposes and compares three different algorithms for this purpose, based on the preliminary extraction of the PWD velocity spectrum envelopes: template matching, supervised classification over a reduced set of relevant waveshape features, and supervised classification over the whole waveshape potentially representing a cardiac cycle. A custom dataset comprising 43 fetal cardiac PWD traces (174,319 signal segments) acquired on an apical five-chamber window was developed and used for the assessment of the different algorithms.
Results:
The adoption of a supervised classifier trained with the samples representing the upper and lower envelopes of the PWD, with additional features extracted from the image, achieved significantly better results (p < 0.0001) than the other algorithms, with an average accuracy of 98% ± 1% when using an SVM classifier and a leave-one-subject-out cross-validation. Further, the robustness of the results with respect to the classifier model was proved.
Conclusions:
The results reveal excellent detection performance, suggesting that the proposed approach can be adopted for the automatic analysis of long PWD traces or embedded in ultrasound machines as a first step for the extraction of measurements and reference clinical parameters.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring pulse
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse
The pulse serves as a clinical...