Halcyon clinical performance evaluation: A log file-based study in comparison with a C-arm Linac.
Ruoxi Wang1, Yi Du1, Kaining Yao1
1Key laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Radiation Oncology, Peking University Cancer Hospital & Institute, Beijing, China.
The Halcyon Treatment Delivery System (TDS) offers comparable mechanical accuracy to the TrueBeam TDS at twice the speed. Dosimetric accuracy, measured by dose-volume parameters, showed smaller variations on the Halcyon, indicating superior performance.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Technology
Background:
- Volumetric Modulated Arc Therapy (VMAT) is a crucial radiotherapy technique.
- Advancements in Treatment Delivery Systems (TDS) aim to improve accuracy and efficiency.
- Comparing new ring-shaped TDS with conventional C-arm linear accelerators (Linacs) is essential for clinical adoption.
Purpose of the Study:
- To compare the dosimetric and mechanical accuracy of VMAT delivery between the Halcyon TDS and a conventional C-arm Linac.
- To evaluate the impact of mechanical parameter fluctuations on dosimetric accuracy.
- To assess the performance of a novel ring-shaped TDS in a clinical setting.
Main Methods:
- Log file analysis of clinically delivered VMAT fractions from Halcyon and TrueBeam Linacs.
- Extraction and analysis of mechanical parameters, including Multi-Leaf Collimator (MLC) leaf errors and gantry angle accuracy.
- Dose recalculations incorporating mechanical parameter fluctuations to evaluate Dose Volume Parameters (DVP).
Main Results:
- Halcyon and TrueBeam demonstrated comparable MLC leaf positioning accuracy (mean RMS errors of 0.028 mm and 0.031 mm, respectively).
- Halcyon exhibited superior gantry angle accuracy (0.035° mean RMS error) compared to TrueBeam (0.065°).
- Both systems showed similar levels of dose-volume parameter variations (approx. 0.1%) due to mechanical errors, but Halcyon had lower DVP standard deviations.
Conclusions:
- The Halcyon TDS achieves mechanical accuracy comparable to the TrueBeam TDS while delivering VMAT at double the speed.
- The Halcyon TDS demonstrates potentially superior dosimetric accuracy, evidenced by lower DVP standard deviations.
- The ring-shaped design of the Halcyon contributes to its enhanced gantry angle precision.