Analytical solution of a personalized intraocular lens design for the correction of spherical aberration and coma of
Bruno Chassagne1, Lionel Canioni2
1ALPhANOV Centre Technologique Optique et Lasers, Institut d'Optique d'Aquitaine, 33400 Talence, France.
Abstract:
This manuscript reports on a closed-form solution determining the personalized required shape of a new intraocular lens able to remove spherical aberration and coma of a pseudophakic eye. The proposed analytical method, within the framework of the Seidel theory of third-order optical aberrations, considers corneal conicities, fourth-order aspheric surface of the intraocular optics, pupil-shift effect and ocular kappa angle.