Use of Artificial Intelligence to Reduce Radiation Exposure at Fluoroscopy-Guided Endoscopic Procedures
Ji Young Bang1, Matthew Hough2, Robert H Hawes1
1Center for Interventional Endoscopy, AdventHealth, Orlando, Florida, USA.
Objectives:
Exposure to ionizing radiation remains a hazard for patients and healthcare providers. We evaluated the utility of an artificial intelligence (AI)-enabled fluoroscopy system to minimize radiation exposure during image-guided endoscopic procedures.
Methods:
We conducted a prospective study of 100 consecutive patients who underwent fluoroscopy-guided endoscopic procedures. Patients underwent interventions using either conventional or AI-equipped fluoroscopy system that uses ultrafast collimation to limit radiation exposure to the region of interest. The main outcome measure was to compare radiation exposure with patients, which was measured by dose area product. Secondary outcome was radiation scatter to endoscopy personnel measured using dosimeter.
Results:
Of 100 patients who underwent procedures using traditional (n = 50) or AI-enabled (n = 50) fluoroscopy systems, there was no significant difference in demographics, body mass index, procedural type, and procedural or fluoroscopy time between the conventional and the AI-enabled fluoroscopy systems. Radiation exposure to patients was lower (median dose area product 2,178 vs 5,708 mGym, P = 0.001) and scatter effect to endoscopy personnel was less (total deep dose equivalent 0.28 vs 0.69 mSv; difference of 59.4%) for AI-enabled fluoroscopy as compared to conventional system. On multivariate linear regression analysis, after adjusting for patient characteristics, procedural/fluoroscopy duration, and type of fluoroscopy system, only AI-equipped fluoroscopy system (coefficient 3,331.9 [95% confidence interval: 1,926.8-4,737.1, P < 0.001) and fluoroscopy duration (coefficient 813.2 [95% confidence interval: 640.5-985.9], P < 0.001) were associated with radiation exposure.
Discussion:
The AI-enabled fluoroscopy system significantly reduces radiation exposure to patients and scatter effect to endoscopy personnel (see Graphical abstract, Supplementary Digital Content, http://links.lww.com/AJG/B461).
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):