Related Experiment Video
Updated: Dec 25, 2025

05:42
Dynamic Navigation for Dental Implant Placement
Published on: September 13, 2022
4.3K
Is there a learning curve in static computer-assisted implant surgery? A prospective clinical study
M Cassetta1, F Altieri1, M Giansanti1
1Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.
International Journal of Oral and Maxillofacial Surgery
|March 29, 2020
Summary
This study investigated the learning curve for static computer-assisted surgery (s-CAS) in dental implantology. Results indicate that a typical learning curve effect was not identified for s-CAS procedures.
Area of Science:
- Dentistry
- Surgical Technology
- Medical Imaging
Background:
- Static computer-assisted surgery (s-CAS) aims to enhance implantology outcomes.
- Understanding the learning curve is crucial for adopting new surgical technologies.
Purpose of the Study:
- To evaluate the learning curve associated with static computer-assisted surgery (s-CAS) in implantology.
- To assess the accuracy of implant placement using s-CAS in partially and totally edentulous patients.
Main Methods:
- Prospective cohort study adhering to STROBE guidelines.
- Two surgeons with implantology experience but no s-CAS experience treated 12 patients (6 partially, 6 totally edentulous).
- Preoperative and postoperative CT scans were used to measure coronal, apical, and angular deviations, and positioning error for 56 implants.
Main Results:
- Mean coronal deviation: 0.87 mm (partially edentulous) and 1.24 mm (totally edentulous).
- Mean apical deviation: 1.13 mm (partially edentulous) and 1.52 mm (totally edentulous).
- Mean angular deviation: 2.63° (partially edentulous) and 3.59° (totally edentulous).
- Mean positioning error: 0.80 mm (partially edentulous) and 1.14 mm (totally edentulous).
Conclusions:
- The study did not identify a typical 'learning curve' effect for static computer-assisted surgery in implantology.
- Surgeons experienced in implantology may adapt to s-CAS without a discernible learning phase based on accuracy metrics.
Keywords:
accuracycomputer-aided implant surgerycomputer-aided implantologycomputer-guided implantologydental implantlearning curvestatic computer-assisted surgery