Validation of a dose tracking software for skin dose map calculation in interventional radiology
Paola Enrica Colombo1, Federica Rottoli1, Marco Felisi1
1Medical Physics Department, ASST Grande Ospedale Metropolitano Niguarda, Piazza Ospedale Maggiore 3, 20162 Milano, Italy.
Summary
The NEXO[DOSE]® software accurately estimates skin dose distribution and Peak Skin Dose (PSD) in interventional radiology (IR) procedures. This tool enables faster and more precise patient monitoring following radiation exposure.
Area of Science:
- Medical Physics
- Radiological Sciences
- Radiation Dosimetry
Background:
- Accurate skin dose monitoring is crucial in interventional radiology (IR) to prevent deterministic radiation effects.
- Existing methods for dose estimation can be complex and time-consuming.
- The NEXO[DOSE]® system aims to provide a streamlined solution for dose assessment in IR.
Purpose of the Study:
- To validate the accuracy of the NEXO[DOSE]® skin dose software for interventional radiology (IR) procedures.
- To assess the software's ability to provide skin dose distribution and Peak Skin Dose (PSD) estimates.
- To compare software calculations with film-based measurements across different angiographic systems.
Main Methods:
- The NEXO[DOSE]® software utilizes radiation dose structured report data and system-specific parameters.
- GafChromic® XR-RV3 films were used as a measurement tool under a PMMA phantom.
- Calculated doses and spatial distributions were compared against film measurements using Philips and Siemens systems.
Main Results:
- Peak Skin Dose (PSD) calculations showed an average agreement of 7% ± 5% with film measurements.
- Discrepancies up to 33% were observed in low-dose regions due to unmodeled scatter.
- Geometric accuracy of dose distribution was within 3-5 mm, demonstrating good spatial agreement.
Conclusions:
- NEXO[DOSE]® software provides a reliable estimation of skin dose distribution and Peak Skin Dose (PSD).
- The software's accuracy supports its use for enhanced patient follow-up in IR.
- Validation across multiple systems indicates consistent performance for dose monitoring.
Keywords:
Interventional RadiologyPeak Skin DoseRadiation Dose Structured ReportRadiochromic filmSkin dose map