Electronic versus conventional spatiotemporal image correlation (STIC) fetal echocardiography: a direct comparison
Spiridon Pepes1, Fragiskos Parthenakis1, Antonis Makrigiannakis2
1School of Medicine, University of Crete, Heraklion, Greece.
Objectives:
Recent advances in Spatial Temporal Image Correlation (STIC) 4 D fetal echocardiography include the application of eSTIC based on electronic probe image acquisition. We aimed to directly compare the performance of conventional STIC versus eSTIC technique (B-Mode and color Doppler imaging) during off-line reconstruction of STIC/eSTIC fetal heart volume pairs.
Methods:
Pairs of B-Mode and Color Doppler STIC volumes were acquired sequentially by firstly conventional (STIC) followed by electronic (eSTIC) probes during 33 consecutive obstetric scans at median 23 (range 13-31) gestational weeks. The resulting 66 fetal heart volume pairs were assessed blindly off-line by a fetal cardiologist who documented feasibility of reconstruction, presence of motion artifacts, subjective image quality on a 4-level scale: 1-best to 4-non-diagnostic and morphological diagnosis, to enable a paired comparison of STIC and eSTIC in the same fetus under similar scanning conditions.
Results:
eSTIC volumes had higher temporal resolution (37 vs. 24 frames per second, p < .001), less motion during acquisition (12 vs. 20 cases, O.R. 7.0, p = .002) and better average image quality (1.9 vs. 2.2, p = .006) compared to STIC volumes. More diagnostic reconstructions were achieved by eSTIC (n = 55, 86%) than STIC (n = 52, 78.8%), p = .001), in a comparable analysis time (mean 4.96 vs. 4.94 min). During a comparison of image quality of the original acquisition (A) and reconstructed planes (B and C planes) e STIC was superior in 22 (33%), 39 (59%) and 21 (38%) volumes, respectively, with the remaining cases being of similar quality (<10% in each plane in favor of STIC). Imaging mode and gestational age had a similar impact on both eSTIC and STIC performance: diagnostically acceptable studies in 49 (75.8%) vs. 48 (72.2%) by B-Mode, 60 (90.9%) vs. 56 (84.8%) by Color Doppler Mode, 8 (62.5%) vs. 10 (50%) in early scans, 38 (95%) vs. 38 (95%) in mid-gestation scans, and 7 (70%) vs. 6 (60%) in third trimester scans. Eight obstetric scans identified a fetus with a cardiac variant or structural abnormality. Diagnostic concordance of the two STIC approaches was comparable (40/48 concordant interpretations, kappa 0.657) all confirmed by fetal and/or postnatal echocardiography.
Conclusions:
eSTIC was associated with more effective 4 D fetal heart reconstruction due to reduced motion artifacts and superior image quality in all planes, when compared to STIC. Early gestation reconstructions were not generally successful using either technology. Further study is needed to define the cost-effectiveness and diagnostic impact of eSTIC over conventional STIC and their role over, or in addition to, screening 2 D fetal echocardiography by appropriately trained sonographers.
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Imaging Studies II: Ultrasonography
Ultrasonography
During an ultrasonography procedure, a handheld device called...