Caffeine Extraction, Enzymatic Activity and Gene Expression of Caffeine Synthase from Plant Cell Suspensions
Published on: October 2, 2018
Extraction-Free, Direct Determination of Caffeine in Microliter Volumes of Beverages by Thermal Desorption-Gas
Xianglu Peng1, Melanie Brown1, Paul Bowdler1
1Department of Applied Sciences, University of the West of England, Frenchay Campus, Coldharbour Lane, Bristol BS16 1QY, UK.
Abstract:
An extraction-free method requiring microliter (μL) volumes has been developed for the determination of caffeine in beverages. Using a pyrolysis-gas chromatography mass spectrometry system, the conditions required for the direct thermal desorption-gas chromatography mass spectrometry (TD-GC/MS) determination of caffeine were optimised. A 5 μL aliquot was introduced to the thermal desorption unit, dried, and thermally desorbed to the GC/MS. The response was linear over the range 10 to 500 μg/mL (R 2 = 0.996). The theoretical limit of detection (3 σ) was 0.456 μg/mL. No interferences were recorded from endogenous beverage components or from commonly occurring drugs, such as nicotine, ibuprofen, and paracetamol. Replicate caffeine determinations on fortified latte style white coffee and Pepsi Max® gave mean recoveries of 93.4% (%CV = 4.1%) and 95.0% (%CV = 0.98%), respectively. Good agreement was also obtained with the stated values of caffeine for an energy drink and for Coca-Cola®. These data suggest that the method holds promise for the determination of caffeine in such samples.
More Related Videos
07:34Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry UPLC-MS
Published on: March 14, 2013
10:13Using Capillary Electrophoresis to Quantify Organic Acids from Plant Tissue: A Test Case Examining Coffea arabica Seeds
Published on: November 12, 2016
Related Concept Videos
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...