Objectives:
Lanthanum (La) retention in bone has been shown to occur in individuals who are orally administered lanthanum carbonate (LaC), a drug to treat hyperphosphatemia. The breakdown of LaC in the gastrointestinal tract into La3+ and carbonate ions results in residual quantities of La being deposited in bone. We previously reported on a non-invasive x-ray fluorescence (XRF) system that was developed to quantify bone La concentrations and applied it to a series of excised cadaver tibiae. However, given interpatient variability in bone shape and size, differential signal attenuation that occurs in bone and tissue, patient movement and overlying tissue thickness at the measurement site, quantifying bone La concentrations during in vivo measurements in live subjects needs to be investigated further along with the radiation dose associated with the measurement.
Approach:
Coherent normalization was investigated as a function of overlying tissue thickness, source-subject distance and bone radius through Monte Carlo simulation and experimental work. This was accomplished by observing the ratio of the net La K x-ray peak area to the coherently scattered peak area at 59.5 keV. In addition, the dose delivered during a 2000 s measurement was determined using radiochromic film.
Main Results:
The coherent normalization of the La x-ray signal was shown to be independent of overlying tissue thickness, source-subject movement and bone radius, which indicates that this normalization procedure can correct for these factors. The equivalent skin dose and effective dose were 18.0 mSv and 3.2 μSv, respectively for a five-year-old.
Significance:
While coherent normalization for the bone lead (Pb) and bone gadolinium (Gd) systems has been shown to be successful, we also report that this normalization procedure can correct for these interpatient variabilities in the in vivo 241Am-La K XRF system.
More Related Videos
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
Related Concept Videos
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...