Myocardial Extracellular Volume Quantification Using Cardiac Computed Tomography: A Comparison of the Dual-energy
Takafumi Emoto1, Seitaro Oda2, Masafumi Kidoh2
1Department of Central Radiology, Kumamoto University Hospital, Chuo-ku, Kumamoto, Japan.
Rationale And Objectives:
To clarify the accuracy of two measurement methods for myocardial extracellular volume (ECV) quantification (ie, the standard subtraction method [ECVsub] and the dual-energy iodine method [ECViodine]) with the use of cardiac CT in comparison to cardiac magnetic resonance imaging (CMR) as a reference standard.
Materials And Methods:
Equilibrium phase cardiac images of 21 patients were acquired with a dual-layer spectral detector CT and CMR, and the images were retrospectively analyzed. CT-ECV was calculated using ECVsub and ECViodine. The correlation between the ECV values measured by each method was assessed. Bland-Altman analysis was used to identify systematic errors and to determine the limits of agreement between the CT-ECV and CMR-ECV values. Root mean squared errors and residual values for the ECVsub and ECViodine were also assessed.
Results:
The correlations between ECVsub and ECViodine for both septal and global measurement were r = 0.95 (p < 0.01) and 0.91 (p < 0.01), respectively, while those between the mean ECVsub and CMR-ECV were r = 0.90 (septal, p < 0.01) and 0.84 (global, p < 0.01), and those between ECViodine and CMR-ECV were r = 0.94 (septal, p < 0.01) and 0.95 (global, p < 0.01). Bland-Altman plots showed lower 95% limits of agreement between ECViodine and CMR-ECV compared with that between ECVsub and CMR-ECV in both septal and global measurement. The root mean squared error of ECVsub was higher than that of ECViodine. The mean residual value of ECVsub was significantly higher than that of ECViodine.
Conclusion:
ECViodine yielded more accurate myocardial ECV quantification than ECVsub, and provided a comparable ECV value to that obtained by CMR.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...