[Creation of Brain Parenchymal Equivalent Phantom for Diffusion Weighted Image Using Microwave Oven: A Different
Maiko Hashimoto1, Masahiko Takahashi1, Norio Hayashi2
1Central Radiology Department, Isesaki Municipal Hospital.
Nihon Hoshasen Gijutsu Gakkai Zasshi
|May 22, 2020
Summary
Creating an acute micro cerebral infarction phantom for MRI is now easier. Using a microwave oven significantly reduces phantom creation time compared to a traditional hot stirrer, with comparable results.
Area of Science:
- Medical Imaging
- Neuroscience
- Biomedical Engineering
Background:
- Acute micro cerebral infarction detection requires specialized phantoms for Magnetic Resonance Imaging (MRI) optimization.
- Traditional phantom creation methods often rely on hot stirrers, which are not widely available in clinical settings.
Purpose of the Study:
- To evaluate the feasibility of using a microwave oven as an alternative to a hot stirrer for creating brain parenchyma phantoms.
- To compare the efficiency and quality of phantoms created using both methods for acute micro cerebral infarction detection.
Main Methods:
- Five acute micro cerebral infarction phantoms were created using a conventional hot stirrer.
- Five phantoms were created using a microwave oven.
- Phantom creation time, T2 values, apparent diffusion coefficient (ADC) values, image uniformity, and reproducibility were assessed and compared between the two methods.
Main Results:
- Microwave oven phantom creation time averaged 108±8 minutes, significantly shorter than the 213±48 minutes required with a hot stirrer.
- No significant differences were observed in T2 values, ADC values, uniformity, or reproducibility between the two phantom creation methods.
Conclusions:
- A microwave oven provides a significantly faster and more accessible method for creating acute micro cerebral infarction phantoms.
- This microwave-assisted technique offers comparable phantom quality to the traditional hot stirrer method, facilitating MRI optimization.