Related Experiment Video
Updated: Dec 19, 2025

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Design of optimal digital filter and digital signal processing for a CdZnTe high resolution gamma-ray system
1Ionizing Radiation Center, Korea Research Institute of Standards and Science (KRISS), Daejeon, 34113, Republic of Korea; University of Science & Technology (UST), Daejeon, 34113, Republic of Korea.
Abstract:
We have developed an online digital signal processing system based on an FPGA. The system consists of pile-up rejection, baseline restorer, peak detection and pole-zero cancellation for evaluation of deposited energy in the detector. The shaping algorithm employed is a Moving Window Deconvolution (MWD) to shape digitized data into a trapezoidal form. For the purpose of verification, the developed digital system was tested for 137Cs gamma rays. The entire system is programmed using the LabVIEW environment.
Related Concept Videos
Properties of the z-Transform I
Design Example