Robotic technology can help compound sterile medications, reducing errors. However, device failures and supply issues limited their ability to meet all daily parenteral dose needs.
Area of Science:
- Pharmacy automation and sterile compounding
- Medical technology and drug delivery systems
Background:
- Compounding sterile medication admixtures is labor-intensive and prone to human error.
- Robotic devices and workflow technology offer potential solutions, particularly for antineoplastic and hazardous medications.
Purpose of the Study:
- To describe single-center experiences with various sterile compounding robotic technologies from 2009 to 2017.
- To analyze the performance and identify barriers associated with robotic compounding of sterile medication admixtures.
Main Methods:
- Evaluation of robotic devices: Intellifill i.v., Cytocare, i.v.Station, i.v.Station ONCO, and i.v.Soft Assist.
- Tracking prepared doses and validating dose accuracy via specific gravity measurement.
- Defining failure criteria: >10% variation for nonhazardous, >5% for hazardous medications.
Main Results:
- Intellifill i.v. compounded over 1,000,000 admixtures (75% of total).
- Other robots compounded 14% (i.v.Station), 7% (Cytocare), 3% (i.v.Station ONCO), and 0.7% (i.v.Soft Assist).
- Common barriers included hardware/software failures and container/supply shortages.
Conclusions:
- Robotic systems show promise but faced significant barriers to optimal performance.
- Device failures, software issues, and supply chain problems limited the ability to meet all daily parenteral dose requirements.