Bitewing radiography dosimetry of a stationary intraoral tomosynthesis imaging system
K Brandon Johnson1, J Colin LaPrade1, Enrique Platin1
1Oral and Maxillofacial Radiology, University of North Carolina Adams School of Dentistry, University of North Carolina at Chapel Hill, NC, USA.
Objectives:
This study assessed effective doses (E) from conventional and stationary intraoral tomosynthesis (s-IOT) radiography for posterior bitewing (PBW) examinations and evaluated the effect of sensor attenuation.
Study Design:
An adult human tissue-equivalent phantom and optically stimulated luminescent dosimeters were used. Series of 4 PBW radiographs were acquired with circular and rectangular collimation. s-IOT PBW radiographs were acquired with built-in rectangular collimation. Radiographs were acquired without and with a sensor in the beam path.
Results:
E (in µSv) was 15.7 and 8.2 for conventional-circular, 4.6 and 1.1 for conventional-rectangular, and 11.9 and 5.9 for s-IOT in sensor-absent and sensor-present scenarios, respectively. For sensor-absent exposures, E for conventional-rectangular was 29.3% and E for s-IOT was 75.8% of the conventional-circular dose. With the sensor present, these values were 13.4% and 72.0%, respectively (P < .001). Sensor-present E was lower than sensor-absent E for all modalities (P < .001). Reductions in equivalent doses were similar to effective dose reductions.
Conclusions:
For PBW examinations, E for s-IOT was smaller than for conventional radiography with circular collimation, but larger than for conventional radiography with rectangular collimation. The presence of a sensor maintained these differences but reduced E for all modalities.