Bitewing radiography dosimetry of the PORTRAY stationary-intraoral tomosynthesis imaging system
K Brandon Johnson1, Elizabeth M Kay2, J Colin LaPrade1
1Department of Oral and Maxillofacial Radiology, UNC Adams School of Dentistry, University of North Carolina-Chapel Hill, Chapel Hill, North Carolina, USA.
Objective:
This study measured effective (E) and equivalent doses from adult and child 3-dimensional (3D) and 2D posterior bitewing (PBW) examinations using the PORTRAY stationary-intraoral tomosynthesis radiography system.
Study Design:
Adult and child phantoms and optically stimulated luminescent dosimeters were used to measure doses for adult-4 and child-2 projection PBW examinations acquired without (W/O) and with (W) a direct digital sensor in the beam path. Child doses without and with thyroid shielding were measured.
Results:
Three-dimensional examination E values (μSv) W/O and W were 16.7 and 7.3 for adult, 9.2 and 3.5 for child, and 8.7 and 3.0 with thyroid shielding, respectively. Two-dimensional examination E values W/O and W were 4.3 and 1.5 for adult, 2.1 and 0.6 for child, and 2.0 and 0.5 with shielding, respectively. Sensor presence reduced E for all adult and child examinations (P = .0001). Child E was reduced compared with adult E for both sensor conditions in 3D (P < .0001) and 2D (P ≤ .0043) imaging. Adult and child 3D W/O and W equivalent thyroid doses did not differ (P ≥ .9996). However, child 2D W/O and W doses were lower (P ≤ .0002). Shielding produced no reduction (P ≥ .1128) for either 3D condition or 2D with the sensor (P = .6615) but reduced child 2D dose without the sensor.
Conclusions:
Inclusion of a sensor yielded significant reductions in adult and child E. Sensor presence impacted thyroid dose reduction more than shielding.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...