Comparison of Two Commercially Available Fentanyl Screening Immunoassays for Clinical Use
Melissa M Budelier1, Caroline E Franks1, Nicole Logsdon1
1Department of Pathology and Immunology, Division of Laboratory and Genomic Medicine, Washington University, St. Louis, MO.
The Journal of Applied Laboratory Medicine
|July 17, 2020
Summary
The ARK fentanyl assay showed better precision and higher concordance with LC-MS/MS than the IAL assay. ARK also detected the fentanyl metabolite norfentanyl, improving overall accuracy in drug testing.
Area of Science:
- Clinical Chemistry
- Forensic Toxicology
- Analytical Chemistry
Background:
- Fentanyl is a potent synthetic opioid linked to widespread illicit drug use and overdose fatalities.
- Two FDA-cleared immunoassays, SEFRIA Immunalysis (IAL) and ARK, are available for fentanyl detection in urine.
- Limited comparative data exists for these two widely used fentanyl detection assays.
Purpose of the Study:
- To directly compare the performance of the IAL and ARK fentanyl immunoassays.
- Evaluate assay precision, cross-reactivity, and concordance with LC-MS/MS.
- Determine the clinical utility of each assay in fentanyl detection.
Main Methods:
- Fentanyl immunoassays (IAL and ARK) were performed on a Roche Cobas e602 analyzer.
- Assay precision was assessed using fentanyl dilutions above, below, and at the 1.0 ng/mL cutoff.
- Cross-reactivity testing included norfentanyl and fentanyl analogs; concordance was evaluated against LC-MS/MS in 90 patient samples.
Main Results:
- ARK assay demonstrated significantly lower imprecision (2.9%-6.4%) compared to IAL (18.9%-40.7%).
- Both assays cross-reacted with acetylfentanyl, acrylfentanyl, and furanylfentanyl; only ARK cross-reacted with norfentanyl.
- ARK assay showed higher concordance with LC-MS/MS (94%) than IAL (90%), improving to 96% when norfentanyl was included in LC-MS/MS confirmation.
Conclusions:
- The ARK fentanyl assay exhibits superior precision and greater concordance with LC-MS/MS compared to the IAL assay.
- ARK's ability to detect norfentanyl enhances its accuracy in identifying fentanyl exposure.
- These findings support ARK as a more reliable immunoassay for fentanyl detection in clinical and forensic settings.