Related Experiment Video
Updated: Dec 14, 2025

Infrared Degenerate Four-wave Mixing with Upconversion Detection for Quantitative Gas Sensing
Published on: March 22, 2019
Partial Least-Squares Regression as a Tool to Retrieve Gas Concentrations in Mixtures Detected Using Quartz-Enhanced
Andrea Zifarelli1,2, Marilena Giglio1,2, Giansergio Menduni2,3
1State Key Laboratory of Quantum Optics and Quantum Optics Devices, Institute of Laser Spectroscopy, Shanxi University, Taiyuan 030006, P. R. China.
Abstract:
We report on a statistical tool based on partial least-squares regression (PLSR) able to retrieve single-component concentrations in a multiple-gas mixture characterized by spectrally overlapping absorption features. Absorption spectra of mixtures of CO-N2O and mixtures of C2H2-CH4-N2O, both diluted in N2, were detected in the mid-IR range by exploiting quartz-enhanced photoacoustic spectroscopy (QEPAS) and using two quantum cascade lasers as light sources. Single-gas reference spectra of each target molecule were acquired and used as PLSR-based algorithm training data set. The concentration range explored in the analysis varies from a few parts-per-million (ppm) to thousands of ppm. Within this concentration range, the influence of the gas matrix on nonradiative relaxation processes can be neglected. Exploiting the ability of PLSR to deal with correlated data, these spectra were used to generate new simulated spectra, i.e., linear combinations of the reference ones. A Gaussian noise distribution was added to the created data set, simulating the real QEPAS signal fluctuations around the peak value. Compared with standard multilinear regression, PLSR predicted gas concentrations with a calibration error up to 5 times better, even with absorption features with spectral overlap greater than 97%.
Related Concept Videos
Atomic Absorption Spectroscopy: Atomization Methods
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Gas Chromatography: Types of Detectors-II
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...