Full Finger Reperfusion Time Measured by Pulse Oximeter Waveform Analysis in Children
Amanda J Nickel1, Shen Jiang2, Natalie Napolitano1
1Department of Respiratory Care, Children's Hospital of Philadelphia, Philadelphia, PA.
Objectives:
Capillary refill time is a noninvasive method to assess tissue perfusion to determine shock status. Capillary refill time is defined as the time required to regain skin color after blanching pressure is applied. Although common methods to measure capillary refill time depend on clinicians' visual assessment, a new approach using a pulse oximeter waveform analysis exists, referred to as full finger reperfusion time. We aim to evaluate reproducibility and validity of the novel full finger reperfusion time measurement using clinicians' visual capillary refill time assessment as a reference standard.
Design:
Prospective observational study.
Setting:
PICUs and operating suites at a large academic children's hospital.
Patients:
Ninety-nine children 1-12 years old with various skin color tones.
Interventions:
Each child had 10 measurements, including five full finger reperfusion time and five clinician capillary refill time, alternating second and third digits.
Measurements And Main Results:
Eighteen children had prolonged capillary refill time (> 2 s) and four children with capillary refill time greater than 3 seconds. Four-hundred eighty-five data pairs were analyzed. Intraclass correlation coefficient of full finger reperfusion time within each patient was 0.76 (95% CI, 0.68-0.83), demonstrating good reproducibility. Correlation coefficient between full finger reperfusion time and clinician capillary refill time was moderate: r = 0.37 (p < 0.0001; 95% CI, 0.29-0.44) for the pairs and r = 0.52 (p < 0.0001; 95% CI, 0.36-0.65) for patient average. Bland-Altman plot showed a consistent difference between full finger reperfusion time and clinician capillary refill time (full finger reperfusion time 1.14 s longer). Weak association was found between force and full finger reperfusion time (β = -0.033 ± 0.016; 95% CI, -0.065 to -0.0016; p = 0.04), finger thickness (β = -0.20 ± 0.089; 95% CI, -0.37 to -0.19; p = 0.03), except for color tone (p = 0.31). Finger temperature was associated with full finger reperfusion time (β = -0.18 ± 0.041; 95% CI, -0.26 to -0.0999; p < 0.0001).
Conclusions:
Full finger reperfusion time demonstrated good reproducibility. Full finger reperfusion time showed moderate correlation with clinician capillary refill time. Full finger reperfusion time was 1.14 seconds longer than capillary refill time. Future studies should focus on the clinical value of full finger reperfusion time as a monitoring device for hemodynamics in critically ill children.
More Related Videos
04:44Author Spotlight: Enhancing Vascular Function and Physical Capacity in Cardiovascular Disease Through Novel Interventions and NIRS Technology
Published on: March 22, 2024
07:27How to Administer Near-Infrared Spectroscopy in Critically ill Neonates, Infants, and Children
Published on: August 19, 2020
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...