Radiomics in diffusion data: a test-retest, inter- and intra-reader DWI phantom study
C Dreher1, T A Kuder2, F König3
1University Hospital Mannheim, Medical Faculty Mannheim, University of Heidelberg, Department of Radiation Oncology, Theodor Kutzer-Ufer 1-3, 68167, Mannheim, Germany; German Cancer Research Center (DKFZ), Division of Radiology, Im Neuenheimer Feld 280, 69120, Heidelberg, Germany.
Aim:
The aim of this study was to evaluate the robustness of radiomics features of a MRI (magnetic resonance imaging) phantom in quantitative diffusion-weighted imaging (DWI) and depending on the image resolution.
Materials And Methods:
Scanning of an in-house developed DWI phantom was performed at a 1.5 T MRI scanner (Magnetom AERA, Siemens, Erlangen, Germany) using an echo planar imaging (EPI) DWI sequence (b=0,500,1,000 s/mm2) with low (3×3 mm2) and high (2×2 mm2) image resolutions. Scans were repeated after phantom repositioning to evaluate retest reliability. Radiomics features were extracted after semi-automatic segmentation and standardised pre-processing. Intra-/interobserver reproducibility and test-retest robustness were assessed using intraclass correlation coefficients (ICC). Differences were tested with non-parametric Wilcoxon's signed-rank and Friedman's test (p < 0.05) with Dunn's post-hoc analysis.
Results:
Test-retest ICC was overall high with >0.90 for 39/46 radiomics features in all sequences/resolutions. Decreased test-retest ICCs were pronounced for conventional Min-value (overall ICC=0.817), and grey-level zone length matrix (GLZLM) features Short-Zone Emphasis (SZE) and Short-Zone Low Grey-level Emphasis (SZLGE) (for both overall ICC=0.927). Test-retest reproducibility was significantly different between b=500, 1,000 and apparent diffusion coefficient (ADC) (mean 0.975±0.050, 0.974±0.051 and 0.966±0.063), which remained significant after post-hoc analysis between b=1,000 and ADC (p = 0.022). ICCs were not significantly different between resolutions of 2×2 and 3×3 mm2 regarding b=500 (mean: 0.977±0.052 and 0.974±0.049, p = 0.612), b=1,000 (mean: 0.973±0.059 and 0.974±0.054, p = 0.516), and ADC (mean: 0.972±0.049 and 0.955±0.101, p = 0.851). Inter- and intra-observer reliability was consistently high for all sequences (overall mean 0.992±0.021 and 0.990±0.028).
Conclusion:
Under ex-vivo conditions, DWI provided robust radiomics features with those from ADC being slightly less robust than from raw DWI (b=500, 1,000 s/mm2). No significant difference was detected for different resolutions. Although, ex-vivo reliability of DWI radiomics features was high, no implications can be made regarding in-vivo analyses.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System IV: CMRI