Related Experiment Video
Updated: Dec 13, 2025

A Microfluidic Model of Biomimetically Breathing Pulmonary Acinar Airways
Published on: May 9, 2016
Flows and function of the Infant Flow neonatal continuous positive airway pressure device investigated with
Thomas Drevhammar1, Niclas Berg2, Kjell Nilsson3
1Department of Women's and Children's Health, Karolinska Institutet, Stockholm, Sweden.
Aim:
The first dedicated neonatal continuous positive airway pressure (CPAP) device using variable flow was the Infant Flow. The system was pressure stable with a low resistance to breathing. The aim of this study was to describe the flow and function of the Infant Flow geometry using simulated breathing and computational fluid dynamics.
Method:
The original Infant Flow geometry was used with simulated term infant breathing at three levels of CPAP. The large eddy simulation methodology was applied in combination with the WALE sub-grid scale model.
Results:
The simulation fully resolved the flow phenomena in the Infant Flow geometry. The main flow feature during inspiration was support by gas entrainment and mixing. During expiration, the jet deflected towards the outlet with unstable impingement of the jet at the opposing edge.
Conclusion:
The proposed mechanism has previously been based on theoretical reasoning, and our results present the first detailed description of the Infant Flow. The pressure stability was based on a jet supporting inspiration by gas entrainment and then being deflected during exhalation. This confirmed previously assumed principles of function and flows within the geometry and provided a base for further developments.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Suctioning the Nasopharyngeal Airway
Equipment Required
Cardiopulmonary Resuscitation II: ACLS Airway Management