Double high-level disinfection versus liquid chemical sterilization for reprocessing of duodenoscopes used for ERCP:
Mark A Gromski1, Marnie S Sieber2, Stuart Sherman1
1Division of Gastroenterology and Hepatology, Department of Medicine, Indiana University School of Medicine, Indianapolis, Indiana, USA.
Background And Aims:
The potential for transmission of pathogenic organisms is a problem inherent to the current reusable duodenoscope design. Recent outbreaks of multidrug-resistant pathogenic organisms transmitted via duodenoscopes has brought to light the urgency of this problem. Microbiologic culturing of duodenoscopes and reprocessing with repeat high-level disinfection (HLD) or liquid chemical sterilization (LCS) have been offered as supplemental measures to enhance duodenoscope reprocessing by the U.S. Food and Drug Administration. This study aims to compare the efficacy of reprocessing duodenoscopes with double HLD (DHLD) versus LCS.
Methods:
We prospectively evaluated 2 different modalities of duodenoscope reprocessing from October 23, 2017 to September 24, 2018. Eligible duodenoscopes were randomly segregated to be reprocessed by either DHLD or LCS. Duodenoscopes were randomly cultured after reprocessing for surveillance based on an internal protocol.
Results:
During the study period, there were 878 post-reprocessing surveillance cultures (453 in the DHLD group and 425 in the LCS group). Of all cultures, 17 were positive for any organism (1.9%). There was no significant difference of positive cultures when comparing the duodenoscopes undergoing DHLD (8 positive cultures, 1.8%) with duodenoscopes undergoing LCS (9 positive cultures, 2.1%; P = .8). Both groups had 2 cultures that grew high-concern organisms (.5% vs .5%, P=1.0). No multidrug-resistant organisms, including carbapenem-resistant enterobacteriaceae, were detected.
Conclusions:
DHLD and LCS both resulted in a low rate of positive cultures, for all organisms and for high-concern organisms. However, neither process completely eliminated positive cultures from duodenoscopes reprocessed with 2 different supplemental reprocessing strategies.
Related Concept Videos
Endoscopic Procedures V: ERCP
Patient...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to: