Related Experiment Video
Updated: Dec 12, 2025

Surgical Bone Implantation Technique for Rat Tibia Models of Diabetes and Osteoporosis
Published on: July 5, 2024
Development of a New Drill Design to Improve the Temperature Control during the Osteotomy for Dental Implants: A
Sergio Alexandre Gehrke1,2, Raphaél Bettach3,4, Benoit Cayron5
1Biotecnos Research Center, Montevideo 11100, Uruguay.
Abstract:
The present in vitro study evaluated a new drill design to improve the temperature control during the osteotomies for dental implant installation, comparing with two drill designs that use conventional external irrigation. Three blocks of synthetic cortical bone were used for osteotomy procedures. Three groups were created: control group 1 (Con1), where a conical multiple drill system with a conventional external irrigation system was used; control group 2 (Con2), where a single bur with a conventional external irrigation system was used; and, test group (Test), where the new single bur (turbo drill) with a new irrigation system was used. Twenty osteotomies were made without irrigation and with intense irrigation, for each group. A thermocouple was used to measure the temperature produced during the osteotomies. The measured temperature were: 28.9 ± 1.68 °C for group Con1; 27.5 ± 1.32 °C for group Con2; 26.3 ± 1.28 °C for group Test. Whereas, the measured temperatures with irrigation were: 23.1 ± 1.27 °C for group Con1; 21.7 ± 1.36 °C for group Con2; 19.4 ± 1.29 °C for group Test. The single drill with a new design for improving the irrigation and temperature control, in comparison with the drill designs with conventional external irrigation.

