Near Infrared Optical Projection Tomography for Assessments of β-cell Mass Distribution in Diabetes Research
Published on: January 12, 2013
Iodine quantification in limited angle tomography
Koen Michielsen1, Alejandro Rodríguez-Ruiz1, Ingrid Reiser2
1Department of Medical Imaging, Radboud University Medical Center, Nijmegen, 6500 HB, The Netherlands.
Purpose:
To develop and test the feasibility of a two-pass iterative reconstruction algorithm with material decomposition designed to obtain quantitative iodine measurements in digital breast tomosynthesis.
Methods:
Contrast-enhanced mammography has shown promise as a cost-effective alternative to magnetic resonance imaging for imaging breast cancer, especially in dense breasts. However, one limitation is the poor quantification of iodine contrast since the true three-dimensional lesion shape cannot be inferred from the two-dimensional (2D) projection. Use of limited angle tomography can potentially overcome this limitation by segmenting the iodine map generated by the first-pass reconstruction using a convolutional neural network, and using this segmentation to restrict the iodine distribution in the second pass of the reconstruction. To evaluate the performance of the algorithms, a set of 2D digital breast phantoms containing targets with varying iodine concentration was used. In each breast phantom, a single simulated lesion with a random size (4 to 8 mm) was placed in a random location within each phantom, with the iodine distribution defined as either homogeneous or rim-enhanced and blood iodine concentration set between 1.4 and 5.6 mg/mL. Limited angle projection data of these phantoms were simulated for wide and narrow angle geometries, and the proposed reconstruction and segmentation algorithms were applied.
Results:
The median Dice similarity coefficient of the segmented masks was 0.975 for the wide angle data and 0.926 for the narrow angle data. Using these segmentations during the second reconstruction pass resulted in an improvement in the concentration estimates (mean estimated-to-true concentration ratio, before and after second pass: 48% to 73% for wide angle; 30% to 73% for narrow angle), and a reduction in the coefficient of variation of the estimates (55% to 27% for wide angle; 54% to 35% for narrow angle).
Conclusion:
We demonstrate that the proposed two-pass reconstruction can potentially improve accuracy and precision of iodine quantification in contrast-enhanced tomosynthesis.
More Related Videos
10:14Synthesis and Purification of Iodoaziridines Involving Quantitative Selection of the Optimal Stationary Phase for Chromatography
Published on: May 16, 2014
07:12Semiautomated Longitudinal Microcomputed Tomography-based Quantitative Structural Analysis of a Nude Rat Osteoporosis-related Vertebral Fracture Model
Published on: September 28, 2017
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Redox Titration: Iodimetry and Iodometry