Related Experiment Video
Updated: Dec 11, 2025

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
A 32-channel receive array coil for bilateral breast imaging and spectroscopy at 7T
Romina Del Bosque1, Jiaming Cui2, Stephen Ogier2
1Department of Biomedical Engineering, Texas A&M University, College Station, Texas, USA.
Purpose:
This work describes the construction and evaluation of a bilateral 32-channel receive array for breast imaging at 7T.
Methods:
The receive array consisted of 32 receive coils, placed on two 3D-printed hemispherical formers. Each side of the receive array consisted of 16 receive loops, each loop having a corresponding detachable board with match/tune capacitors, active detuning circuitry, and a balun. Coil performance was evaluated on homogeneous canola oil phantoms using a Philips Achieva 7T system. Array coil performance was compared with a bilateral forced current excitation volume coil in transmit/receive mode and with a previously reported 16-channel unilateral coil with a similar design.
Results:
The 32-channel array had an increase in average SNR throughout both phantoms by a factor of five as compared with the volume coil, with SNR increases up to 10 times along the periphery and three times in the center. Noise measurements showed low interelement noise correlation (average: 5.4%; maximum: 16.8%). Geometry factor maps were acquired for various acceleration factors and showed mean geometry factors <1.2, for combined acceleration factors of up to six.
Conclusions:
The improvements achieved demonstrate the clear potential for use in dynamic contrast-enhanced or diffusion-weighted MR studies, while maintaining diagnostically relevant spatial and temporal resolutions.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies II: Ultrasonography
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System IV: CMRI
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...