Calibration Transfer of Partial Least Squares Regression Models between Desktop Nuclear Magnetic Resonance
Diego Galvan1, Evandro Bona2, Dionisio Borsato1
1Departamento de Química, Universidade Estadual de Londrina, 86.057-970 Londrina, Brazil.
Abstract:
Low-field proton nuclear magnetic resonance (LF-1H NMR) devices based on permanent magnets are a promising analytical tool to be extensively applied to the process analytical chemistry scenario. To enhance its analytical applicability in samples where the spectral resolution is compromised, multivariate regression methods are required. However, building a robust calibration model, such as partial least squares (PLS) regression, is a laborious task because (1) the number of measurements required during the calibration process is large and (2) the procedure must be repeated when the instrument is changed or after a certain period due to the long-term stability of the instrument. Thus, the present work describes the application of calibration transfer methodologies (direct standardization (DS), piece-wise direct standardization (PDS), and double-window piece-wise direct standardization (DWPDS)) on LF-1H NMR to exempt the necessity of a recalibration procedure when moving from the original spectrometer to a second one with the same, lower, or higher magnetic field. These calibration transfer methodologies were tested with PLS models built on a 60 MHz (for the proton Larmor frequency) spectrometer to predict the specific gravity (SG), distillation temperature (T50%), and final boiling point (FBP) of commercial gasoline. The results showed that the DWPDS method applying only 2 to 7 transference samples enables the transference of all PLS models built on the primary instrument (60 MHz) to other (43, 60, and 80 MHz) different instruments, reaching the same RMSEP values as the primary instrument: 1.2 kg/m3 for SG, 5.1 °C for FBP, and 1.1 °C for T50%.
More Related Videos
10:22Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
07:24Paramagnetic Relaxation Enhancement for Detecting and Characterizing Self-Associations of Intrinsically Disordered Proteins
Published on: September 23, 2021
Related Concept Videos
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
NMR Spectrometers: Resolution and Error Correction
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
NMR Spectrometers: Overview
Calibration Curves: Correlation Coefficient
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...