Abstract:
A CE instrument that can be assembled from commercially available components with minimal construction effort is described. Except for the electronic control circuitry no specially made parts are required. It is based on a flexible design of microfluidic, electropneumatic, and electronic sections and different configurations can easily be implemented. Automated injection into the capillary is performed hydrodynamically by the application of a pressure for a controlled length of time. The performance of the device was tested with a contactless conductivity detector by separating different metal ions. In addition, nine metal cations related to the quality of honey were separated in 2.3 min and four honey samples were analysed quantitatively to demonstrate the applicability of the method.