A Solution for Homogeneous Liver Enhancement in Computed Tomography: Results From the COMpLEx Trial
Bibi Martens, Joachim E Wildberger, Babs M F Hendriks
1Department of Clinical Epidemiology and Medical Technology Assessment, Maastricht University Medical Center.
Objectives:
The aim of the study was to reach homogeneous enhancement of the liver, irrespective of total body weight (TBW) or tube voltage. An easy-to-use rule of thumb, the 10-to-10 rule, which pairs a 10 kV reduction in tube voltage with a 10% decrease in contrast media (CM) dose, was evaluated.
Materials And Methods:
A total of 256 patients scheduled for an abdominal CT in portal venous phase were randomly allocated to 1 of 4 groups. In group 1 (n = 64), a tube voltage of 120 kV and a TBW-adapted CM injection protocol was used: 0.521 g I/kg. In group 2 (n = 63), tube voltage was 90 kV and the TBW-adapted CM dosing factor remained 0.521 g I/kg. In group 3 (n = 63), tube voltage was reduced by 20 kV and CM dosing factor by 20% compared with group 1, in line with the 10-to-10 rule (100 kV; 0.417 g I/kg). In group 4 (n = 66), tube voltage was decreased by 30 kV paired with a 30% decrease in CM dosing factor compared with group 1, in line with the 10-to-10 rule (90 kV; 0.365 g I/kg). Objective image quality was evaluated by measuring attenuation in Hounsfield units (HU), signal-to-noise ratio, and contrast-to-noise ratio in the liver. Overall subjective image quality was assessed by 2 experienced readers by using a 5-point Likert scale. Two-sided P values below 0.05 were considered significant.
Results:
Mean attenuation values in groups 1, 3, and 4 were comparable (118.2 ± 10.0, 117.6 ± 13.9, 117.3 ± 21.6 HU, respectively), whereas attenuation in group 2 (141.0 ± 18.2 HU) was significantly higher than all other groups (P < 0.01). No significant difference in attenuation was found between weight categories 80 kg or less and greater than 80 kg within the 4 groups (P ≥ 0.371). No significant differences in subjective image quality were found (P = 0.180).
Conclusions:
The proposed 10-to-10 rule is an easily reproducible method resulting in similar enhancement in portal venous CT of the liver throughout the patient population, irrespective of TBW or tube voltage.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
09:23Author Spotlight: Enhancing Transplantation Research Through MicroCT Angiography in Murine Models
Published on: September 22, 2023
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...