Optical Performance of a Monofocal Intraocular Lens Designed to Extend Depth of Focus
Purpose:
To test the performance of a new monofocal intraocular lens, intended to extend depth of focus (Tecnis Eyhance, ICB00; Johnson & Johnson Vision, Inc) (ICB-IOL), in comparison to a standard monofocal IOL (Tecnis 1-piece, ZCB00; Johnson & Johnson Vision, Inc) (ZCB-IOL) of the same platform and material.
Methods:
Assessment of the optical performance of the two IOLs was made in vitro using an optical test bench with a model eye. The spherical aberration, modulation transfer function (MTF), and area under the MTF (MTFa) were obtained for pupil sizes ranging from 2 to 5 mm. Through-focus MTFa curves between -3.00 and +1.00 diopters (D) were obtained with three pupil sizes (2, 3, and 4.5 mm). Halo formation was also assessed for both lenses.
Results:
The ICB-IOL had slightly worse optical quality at its best focus (ie, lower MTF scores at distance vision) and more negative spherical aberration than the ZCB-IOL for pupils ranging from 2 to 3 mm. The maximum of the through-focus MTFa curve of the ICB-IOL with a 2-mm pupil shifted to a myopic defocus of -0.50 D. For larger pupils (≥ 3.5 mm), there were no differences in spherical aberration, MTF scores, and halo energy between the two lenses.
Conclusions:
The new ICB-IOL is a modified monofocal lens with 0.50 D of additional power in its central 2-mm zone and more negative spherical aberration values, which induce a myopic shift of the maximum of optical quality and could improve intermediate vision. For pupils larger than 3.5 mm, there were no differences between IOLs. The new ICB-IOL design would produce photic phenomena comparable to a standard IOL. [J Refract Surg. 2020;36(9):625-632.].
More Related Videos
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
05:14Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Related Concept Videos
Focusing of Light in the Eye
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Depth Perception and Spatial Vision
Confocal Fluorescence Microscopy