Purpose:
Mechanical heart valves (MHVs) are widely used to replace diseased heart valves, but it may suffer from cavitation due to the rapid closing velocity of the leaflets, resulting in the damage of red blood cells and platelets. The aim of this study is to apply computational fluid dynamics (CFD) method to investigate the cavitation in bileaflets mechanical heart valves (BMHVs) and discuss the effects of the conduit and leaflet geometries on cavitation intensity.
Methods:
Firstly, CFD method together with moving-grid technology were applied and validated by comparing with experimental results obtained from other literature. Then the leaflets movement and the flow rate of BMHVs with different conduit geometries and leaflet geometries are compared. At last, the duration time of the saturated vapor pressure and the closing velocity of leaflets at the instant of valve closure were used to represent the cavitation intensity.
Results:
Larger closing velocity of leaflets at the instant of valve closure means higher cavitation intensity. For BMHVs with different conduit geometries, the conduit with Valsalva sinuses has the maximum cavitation intensity and the straight conduit has the minimum cavitation intensity, but the leaflets cannot reach the fully opened state in a straight conduit. For BMHVs with different leaflet geometries, in order to minimize the cavitation intensity, the leaflets are better to have a large thickness and a small rotational radius.
Conclusion:
CFD method is a promising method to deal with cavitation in BMHVs, and the closing velocity of leaflets has the same trend with the cavitation intensity. By using CFD method, the effects of the conduit geometry and the leaflet geometry on cavitaion in BMHVs are found out.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Regurgitation III: Medical Management
Aortic Regurgitation III: Medical Management
Mitral Stenosis III: Medical Management
Mitral Regurgitation I: Introduction
Aortic Regurgitation I: Introduction