Effect of the feed injection method on band broadening in analytical supercritical fluid chromatography
Kim Vanderlinden1, Gert Desmet1, Ken Broeckhoven1
1Vrije Universiteit Brussel, Department of Chemical Engineering, Pleinlaan 2, 1050 Brussels, Belgium.
Abstract:
The behavior of a novel type of SFC injector, the feed injector, was investigated. In SFC, the sample compounds are usually diluted in a solvent which has a higher elution strength than the mobile phase, which leads to solvent mismatch upon injection and evidently band broadening. The feed injector differs from standard injectors as the sample, contained in the sample needle or loop, is not switched in line with the mobile phase flow, but directly injected/added to the mobile phase flow (F). The subsequent mixing of sample and mobile phase flows inherently results in a dilution of the sample, thus reducing the solvent mismatch. However, for a given injection/feed flow rate Ffeed, the total volume in which the sample is contained increases with a factor (Ffeed + F)/Ffeed. In addition, to ensure that all of the loaded sample is injected on the column, an additional overfeed volume (Vov) needs to be injected after the sample plug. To better understand the effect of these operating parameters, a wide range of injection conditions was investigated by varying the Ffeed/F-ratio, Vov, overfeed solvent etc. under SFC conditions. It was found that an optimal Ffeed/F exists which is independent of F and decreases with increasing solvent strength dependency of the sample compound. Decreasing Vov has a beneficial effect on peak dispersion but can only be varied over a certain range to ensure the full injection of the loaded sample. On the other hand, it was found that a much larger gain could be made by switching the overfeed solvent to one more compatible with the CO2-based mobile phase. Further reduction of the band broadening could be achieved by applying partial sample injections.
More Related Videos
10:41Ion Exchange Chromatography IEX Coupled to Multi-angle Light Scattering MALS for Protein Separation and Characterization
Published on: April 5, 2019
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...