Validation of a New Transcutaneous tcPO2/tcPCO2 Sensor with an Optical Oxygen Measurement in Preterm Neonates
Willem van Weteringen1,2, Tanja van Essen3, Norani H Gangaram-Panday3
1Department of Pediatric Surgery, Erasmus MC - Sophia Children's Hospital, University Medical Center Rotterdam, Rotterdam, The Netherlands, w.vanweteringen@erasmusmc.nl.
Introduction:
Traditional transcutaneous oxygen (tcPO2) measurements are affected by measurement drift, limiting accuracy and usability. The new potentially drift-free oxygen fluorescence quenching technique has been combined in a single sensor with conventional transcutaneous carbon dioxide (tcPCO2) monitoring. This study aimed to validate optical tcPO2 and conventional tcPCO2 against arterial blood gas samples in preterm neonates and determine measurement drift.
Methods:
In this prospective observational study, during regular care, transcutaneous measurements were paired to arterial blood gases from preterm neonates aged 24-31 weeks of gestational age (GA) with an arterial catheter. Samples were included based on stability criteria and stratified for sepsis status. Agreement was assessed using the Bland-Altman analysis. Measurement drift per hour was calculated.
Results:
Sixty-eight premature neonates were included {median (interquartile range [IQR]) GA of 26 4/7 [25 3/7-27 5/7] weeks}, resulting in 216 stable paired samples. Agreement of stable samples in neonates without sepsis (n = 38) and with suspected sepsis (n = 112) was acceptable for tcPO2 and good for tcPCO2. However, in stable samples of neonates with sepsis (n = 66), tcPO2 agreement (bias and 95% limits of agreement) was -32.6 (-97.0 to 31.8) mm Hg and tcPCO2 agreement was 4.2 (-10.5 to 18.9) mm Hg. The median (IQR) absolute drift values were 0.058 (0.0231-0.1013) mm Hg/h for tcPO2 and 0.30 (0.11-0.64) mm Hg/h for tcPCO2.
Conclusion:
The accuracy of optical tcPO2 in premature neonates was acceptable without sepsis, while electrochemically measured tcPCO2 remained accurate under all circumstances. Measurement drift was negligible for tcPO2 and highly acceptable for tcPCO2.
More Related Videos
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...