Clinical evaluation of an application aid for less-invasive surfactant administration (LISA)
Christian Achim Maiwald1,2, Patrick Neuberger3, Axel R Franz4,2
1Department of Neonatology, University Children's Hospital Tubingen, Tubingen, Germany Christian.Maiwald@med.uni-tuebingen.de.
Background:
Less-invasive surfactant administration (LISA) is increasingly used. We investigated the feasibility of a new LISA-device (Neofact®) in neonates.
Design:
Prospective observational pilot study with open-label LISA in two tertiary neonatal intensive care units.
Patients:
20 infants with a gestational age of ≥26+0/7 weeks and an indication for LISA (Respiratory Severity Score (RSS)≥5 or fraction of inspired oxygen (FiO2) ≥0.30). Infants with respiratory tract malformations or unavailability of an instructed neonatologist were excluded.
Main Outcome Measures:
Success of LISA, defined as laryngoscopy-confirmed intratracheal catheter position or a decrease in FiO2 by ≥0.05 or to 0.21, accompanied by an RSS decrease of ≥2; number of attempts needed for tracheal catheterisation.
Results:
20/57 screened infants were enrolled. Successful application occurred in 19/20 (95%). One application failed after three attempts. No device-related adverse events occurred. The median number of attempts was 2, success rate per attempt 19/31 (61%).
Conclusion:
LISA via Neofact® appears feasible.