Objectives:
The aim of the current study was to evaluate potential differences in the accuracy of mandibular reconstruction and long-term stability, with respect to different reconstructive procedures.
Methods:
In total, 42 patients who had undergone primary segmental mandibular resection with immediate alloplastic reconstruction, with either manually pre-bent or patient-specific mandibular reconstruction plates (PSMRP), were included in this study. Mandibular dimensions, in terms of six clinically relevant distances (capitulum [most lateral points], capitulum [most medial points], incisura [most caudal points], mandibular foramina, coronoid process [most cranial points], dorsal tip of the mandible closest to the gonion point) determined from tomographic images, were compared prior to, and after surgery.
Results:
Dimensional alterations were significantly more often found when conventionally bent titanium reconstruction plates were used. These occurred in the area of the coronoid process (p = 0.014). Plate fractures were significantly (p = 0.022) more often found within the manually pre-bent group than within the PSMRP group (17%/0%).
Conclusion:
The results suggest that the use of PSMRP may prevent rotation of the proximal mandibular segment, thus avoiding functional impairment. In addition, the use of PSMRP may potentially enhance the long-term stability of alloplastic reconstructions.