Liquid Chromatography-Tandem Mass Spectrometry-Based α1-Antitrypsin (AAT) Testing
Josiah D Murray1, Maria A Willrich1, Michael J Krowka2
1Department of Laboratory Medicine and Pathology.
Objectives:
Failure to produce sufficient quantities of functional α1-antitrypsin (AAT) can result in AAT deficiency (AATD) and significant comorbidities. Laboratory testing plays a vital role in AATD, with diagnosis requiring documentation of both a low AAT level and a mutated allele. This retrospective evaluation examines the efficacy of a liquid chromatography-tandem mass spectrometry (LC-MS/MS) (proteotyping)-based algorithm for AATD detection.
Methods:
A 16-month retrospective data analysis was performed on two cohorts: 5,474 samples tested with the proteotype-based algorithm and 16,147 samples directly tested by isoelectric focusing (IEF) phenotyping.
Results:
LC-MS/MS reduced the rate of IEF testing by 97%. The 3% of cases reflexed to IEF resulted in 12 (0.2%) additional phenotype findings. Retrospectively applying the proteotype-based algorithm to the IEF cohort demonstrated a 99.9% sensitivity for the detection of deficiency-associated phenotypes. Most deficiency phenotypes missed by the proteotyping algorithm would come from heterozygous patients with an F, I, or P paired to an S or Z. In all of these cases, patient AAT levels were greater than 70 mg/dL, above the threshold for AAT augmentation therapy.
Conclusions:
The proteotype algorithm is a sensitive and cost-effective approach for the diagnosis of clinical AAT deficiency.
More Related Videos
09:36Author Spotlight: Discovering New Alkaloids in Plants with Advanced Mass Spectrometry Techniques
Published on: March 8, 2024
08:40Millisecond Hydrogen/Deuterium-Exchange Mass Spectrometry for the Study of Alpha-Synuclein Structural Dynamics Under Physiological Conditions
Published on: June 23, 2022
Related Concept Videos
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
MALDI-TOF Mass Spectrometry
Matrix-assisted laser desorption ionization (MALDI) is a commonly...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...