Related Experiment Video
Updated: Dec 4, 2025

Automated 3D Optical Coherence Tomography to Elucidate Biofilm Morphogenesis Over Large Spatial Scales
Published on: August 21, 2019
Comparison of central topographic maps from a swept-source OCT biometer and a Placido disk-dual Scheimpflug
Li Wang1, Ana Laura C Canedo, Yvonne Wang
1From the Department of Ophthalmology, Cullen Eye Institute, Baylor College of Medicine, Houston, Texas, USA.
Purpose:
To compare central topography (CT) from IOLMaster 700 with predicate topographic (PT) maps from a Placido disk-dual Scheimpflug tomographer in detection of irregularities that would influence the decision-making for implanting premium intraocular lenses (IOLs) (toric, multifocal, or extended depth-of-focus).
Setting:
Cullen Eye Institute, Baylor College of Medicine, Houston, Texas.
Design:
Prospective comparative case series.
Methods:
Eyes with various corneal conditions were randomly selected: regular/irregular corneas, previous corneal refractive surgery, and keratoconus or pellucid marginal degeneration. Three observers compared the CT and PT maps and answered a questionnaire for each eye. The questionnaire focused on (1) overall shape similarity between CT and PT and (2) decision-making for premium IOL implantation based on CT and PT. Answers to the questionnaire and agreement in answers among observers were evaluated.
Results:
The study included 105 eyes. Comparing CT and PT, similar shape was observed in 68.6% to 89.5% of cases, and comparable map symmetries were reported in 60.0% to 83.8%; the same decision regarding premium IOL implantation was made in 75.2% to 97.1% of cases. There were significant interobserver agreements among 3 observers for all questions, with Fleiss κ values ranging from 0.141 to 0.450 (all P < .05). Peripheral corneal steeping or flattening was the primary finding that was visible on PT but not on CT.
Conclusions:
Compared with the PT, CT provided similar overall shape and comparable symmetries in most cases. The same decision was made whether to recommend a premium IOL based on CT and PT in 75% to 97% of cases.
Related Concept Videos
Methods of Obtaining Topography
Plotting of Topographic Maps
Topographic Surveying and Contours
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...