Pulmonary hypertensive crisis: A potential reason for right ventricle and pacemaker lead failure
Merve Maze Aydemir1, Hasan Candas Kafali1, Hakan Gemici2
1Department of Pediatric Cardiology, Istanbul Mehmet Akif Ersoy Thoracic and Cardiovascular Surgery Training and Research Hospital, University of Health Sciences, Istanbul, Turkey.
Abstract:
Cardiac pacemakers have improved patient survival and quality of life, although malfunctions can be seen. We present the case of a girl with Seckel syndrome and congenital complete heart block. She had a single chamber permanent pacemaker in the right ventricle. When she referred us with a pulmonary hypertensive crisis (PHC), it was seen that the device was not pacing even in maximum threshold and pulse width values. After new epicardial lead implantation into the left ventricular apex, capture could be established again. For the cases presenting with capture failure, after eliminating lead-related problems and biochemical abnormalities, PHC should be kept in mind as a reason.
More Related Videos
07:41Induction and Phenotyping of Acute Right Heart Failure in a Large Animal Model of Chronic Thromboembolic Pulmonary Hypertension
Published on: March 17, 2022
07:41Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
Related Concept Videos
Mitral Regurgitation I: Introduction
Mitral Stenosis I: Introduction
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Cardiac Catheterization II: Right Heart Catheterization
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy III: Hypertrophic Cardiomyopathy