Surface guided imaging during stereotactic radiosurgery with automated delivery
Elizabeth L Covington1, Dennis N Stanley1, John B Fiveash1
1Department of Radiation Oncology, University of Alabama - Birmingham, Birmingham, AL, USA.
Journal of Applied Clinical Medical Physics
|October 23, 2020
Summary
Surface imaging effectively monitors patient motion during automated intracranial radiotherapy. This method offers a reliable alternative to traditional imaging, ensuring treatment accuracy with minimal disruption.
Area of Science:
- Medical Physics
- Radiation Oncology
- Image Guidance
Background:
- Frameless intracranial stereotactic radiotherapy requires precise patient immobilization.
- Monitoring intrafraction motion is crucial for treatment accuracy.
- Automated delivery systems like HyperArc aim to streamline radiotherapy delivery.
Purpose of the Study:
- To evaluate the efficacy of surface-guided imaging (SI) for monitoring intrafraction motion.
- To assess the performance of the AlignRT® system during frameless intracranial stereotactic radiotherapy with HyperArc™ automated delivery.
- To investigate potential influences of patient demographics and treatment parameters on motion monitoring accuracy.
Main Methods:
- Utilized the AlignRT® surface imaging system to monitor intrafraction motion in patients undergoing HyperArc™ radiotherapy.
- Correlated AlignRT® log files with treatment trajectory logs to analyze motion offsets.
- Investigated performance at non-zero couch angles and during camera-pod blockage.
- Analyzed motion data stratified by patient race to assess potential skin tone-related differences.
Main Results:
- Median reported motion from treatment start to end was 0.24 mm.
- Median offset before beam delivery at non-zero couch angles was 0.55 mm.
- Median motion magnitude during gantry motion with camera-pod blockage was below 1 mm.
- No significant differences in motion offsets were observed for patients stratified by race.
Conclusions:
- Surface image guidance is a practical and effective method for monitoring intrafraction motion.
- SI provides a viable alternative to scheduled mid-treatment imaging during automated stereotactic radiosurgery.
- The AlignRT® system demonstrates reliable performance in conjunction with HyperArc™ for frameless intracranial radiotherapy.