The nano-scale viscoelasticity using atomic force microscopy in liquid environment
Shatruhan Singh Rajput1, Surya Pratap S Deopa1, Jyoti Yadav2
1Department of Physics, Indian Institute of Science Education and Research Pune, Dr. Homi Bhabha Road, Pashan, Pune 411008, India.
Nanotechnology
|October 29, 2020
Summary
This study reveals that single protein molecules exhibit immeasurably low dissipation, challenging previous findings. Atomic force microscopy measurements highlight potential artifacts in conventional methods for quantifying nanoscale viscoelasticity.
Area of Science:
- Nanoscale science
- Biophysics
- Physical chemistry
Background:
- Quantifying viscoelasticity in nanoscale systems is crucial for understanding molecular behavior.
- Previous studies reported significant dissipation in single protein molecules, but experimental limitations may have contributed to these findings.
- Atomic force microscopy (AFM) is a key technique for probing mechanical properties at the nanoscale.
Purpose of the Study:
- To accurately measure the viscoelasticity of single protein molecules and confined water layers.
- To investigate and identify sources of artifacts in AFM-based dissipation measurements.
- To compare the performance of different AFM detection schemes for viscoelasticity measurements.
Main Methods:
- Utilized two AFM detection schemes: commercial deflection detection (cantilever bending) and fiber-interferometer based detection (cantilever displacement).
- Modeled cantilever hydrodynamics using the Euler-Bernoulli equation to account for different detection methods.
- Performed measurements on single octomers of titin (I27 b8) and molecular layers of water confined between solid surfaces.
Main Results:
- The dissipation coefficient of single titin I27 b8 was found to be immeasurably low (upper bound 5 × 10 b-7 kg s b-1), contradicting literature values.
- Entropic stiffness of single unfolded protein domains was measured at approximately 10 mN m b-1 using both methods.
- Identified phase artifacts in conventional deflection detection as a major source of error in dissipation estimates.
Conclusions:
- Cantilever displacement measurement offers superior accuracy for artifact-free viscoelasticity measurements compared to bending detection, especially at low frequencies.
- The low dissipation observed in single protein molecules suggests a need to re-evaluate their role in energy dissipation mechanisms.
- Findings provide insights into discrepancies in protein collapse dynamics observed between force spectroscopy and optical techniques.