BENCHMARKING OF A NEW AUTOMATIC CT RADIATION DOSE CALCULATOR
Kofi Okyere Akyea-Larbi1,2, Mercy Afadzi Tetteh2,3, Anne Catrine T Martinsen3,4
1Department of Medical Physics, School of Nuclear and Allied Sciences, University of Ghana, LG 25 Legon, Accra, Ghana.
Abstract:
Information on patient radiation dose is essential to meet the radiation protection regulations and the demands of dose optimization. Vendors have developed different tools for patient dose assessment for radiological purposes. In this study, estimated effective doses derived from a new image-based software tool (DoseWatch, GE Healthcare) was benchmarked against the corresponding doses from a dose calculator (CT-Expo, SASCRAD) and a conversion coefficient method. Dose data from 150 adult patients (66 male and 84 female), who underwent CT head, abdominopelvic or chest examinations, were retrospectively collected using DoseWatch. Effective dose estimated by DoseWatch was significantly lower than that of CT-Expo and DLP-E (k) (p ≤ 0.001). For the organ doses, DoseWatch resulted in lower dose than CT-Expo for all the organs with the exception of testis (p ≤ 001) and eye lenses (p ≤ 0.026).