Related Experiment Video
Updated: Nov 29, 2025

Author Spotlight: Introduction to Active Probe Atomic Force Microscopy with Quattro-Parallel Cantilever Arrays
Published on: June 13, 2023
2D Scanning Micromirror with Large Scan Angle and Monolithically Integrated Angle Sensors Based on Piezoelectric Thin
Katja Meinel1, Marcel Melzer1, Chris Stoeckel1,2
1Center for Microtechnologies, Chemnitz University of Technology, 09111 Chemnitz, Germany.
Abstract:
A 2D scanning micromirror with piezoelectric thin film aluminum nitride (AlN), separately used as actuator and sensor material, is presented. For endoscopic applications, such as fluorescence microscopy, the devices have a mirror plate diameter of 0.7 mm with a 4 mm2 chip footprint. After an initial design optimization procedure, two micromirror designs were realized. Different spring parameters for x- and y-tilt were chosen to generate spiral (Design 1) or Lissajous (Design 2) scan patterns. An additional layout, with integrated tilt angle sensors, was introduced (Design 1-S) to enable a closed-loop control. The micromirror devices were monolithically fabricated in 150 mm silicon-on-insulator (SOI) technology. Si (111) was used as the device silicon layer to support a high C-axis oriented growth of AlN. The fabricated micromirror devices were characterized in terms of their scanning and sensor characteristics in air. A scan angle of 91.2° was reached for Design 1 at 13 834 Hz and 50 V. For Design 2 a scan angle of 92.4° at 12 060 Hz, and 123.9° at 13 145 Hz, was reached at 50 V for the x- and y-axis, respectively. The desired 2D scan patterns were successfully generated. A sensor angle sensitivity of 1.9 pC/° was achieved.