All-Dielectric Metasurface Fluorescence Biosensors for High-Sensitivity Antibody/Antigen Detection
1Research Center for Functional Materials, National Institute for Materials Science (NIMS), 1-1 Namiki, Tsukuba 305-0044, Japan.
Abstract:
Protein biomolecules are used as markers for various diseases and infections and are targets in biosensing research and development. One of the highest-sensitivity biosensing techniques is considered to be fluorescence (FL) sensing. However, to our knowledge, no study has shown that all-dielectric metasurfaces can contribute to highly sensitive FL sensing. Here, we introduce an efficient type of FL-sensing platforms and show that all-dielectric metasurface FL biosensors are able to directly detect a representative antibody, immunoglobulin G, at very small concentrations on the order of pg/mL or tens of femtomolars. Furthermore, it is shown that they work efficiently as an indirect detection platform for standard cancer marker antigens, such as carcinoembryonic antigens, at concentrations well below the medical diagnosis criterion at 5 ng/mL. Importantly, the metasurface biosensors simultaneously suppress inhomogeneous FL responses, exhibit high reproducibility, and retain sensitivity, even in human serum. These results indicate that the present metasurface FL biosensors provide a high-sensitivity practical platform, suggesting that they are a better option than the commercially standard of enzyme-linked immunosorbent assays.