Dealing with RV-oversensing; separate sensitivity settings for brady and tachy sensing
Pranav Bhagirath1, Kyle Beunder1, Vokko van Halm1
1Department of Cardiology, Amsterdam University Medical Center, Amsterdam, The Netherlands.
Abstract:
An 81-year-old male with a history of systolic heart failure due to an underlying ischemic cardiomyopathy with a left ventricular ejection fraction of 13% and QRS duration of 130 ms had undergone an uncomplicated cardiac resynchronization therapy defibrillator implantation (Quadra Assura MP, St. Jude Medical, LV lead (SJM Quartet 1458Q-86), RA lead (Biotronik Safio S53) and RV shocklead (Biotronik Linox Smart S65 ProMRI) in 2015.
More Related Videos
09:56Implantation of Combined Telemetric ECG and Blood Pressure Transmitters to Determine Spontaneous Baroreflex Sensitivity in Conscious Mice
Published on: February 14, 2021
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Decreased pulse rate
There are specific risk factors that can elevate the likelihood of developing bradycardia. Advanced age is a significant factor, with...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias II: Classification of Tachyarrhythmias
Increased pulse rate
Many factors can elevate the risk of developing tachycardia. These include advanced age, a family history of arrhythmias, and an...
Dysrhythmias VI: Management of Dysrhythmias
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...