Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Repeatability and reproducibility of manifest refraction
Suphi Taneri1, Samuel Arba-Mosquera, Anika Rost
1From the Center for Refractive Surgery, Eye Department at St. Francis Hospital (Taneri, Rost, Kießler), Muenster, Ruhr-University, University Eye Hospital (Taneri, Dick), Bochum, and SCHWIND eye-tech-solutions (Arba-Mosquera), Kleinostheim, Germany.
Manifest refraction repeatability and reproducibility were evaluated, showing high consistency. This suggests that multiple preoperative refractions can improve refractive surgery outcomes and reduce the need for enhancements.
Area of Science:
- Ophthalmology
- Optometry
- Vision Science
Background:
- Accurate refractive measurements are crucial for successful refractive surgery.
- Understanding the consistency of manifest refraction is essential for patient outcomes.
Purpose of the Study:
- To assess the intraexaminer repeatability and interobserver reproducibility of manifest refraction.
- To determine the reliability of refractive measurements in a clinical setting.
Main Methods:
- Retrospective study at a tertiary referral center.
- Analysis of manifest refractions from 1000 eyes of patients undergoing refractive surgery.
- Standardized protocol using automated phoropter, wavefront, tomography, and topography.
Main Results:
- Mean pairwise absolute difference in spherical equivalent was 0.16 ± 0.19 diopters.
- 95% limits of agreement were within 0.50 D for sphere, cylinder, and spherical equivalent.
- Astigmatism axis reproducibility was approximately 10 degrees (SD), with 95% LoA within 22 degrees.
Conclusions:
- Manifest refraction demonstrated high repeatability and reproducibility, with a mean difference of 0.16 D.
- No significant differences were found between repeatability and reproducibility for key refractive parameters.
- Preoperative multiple refractions may enhance surgical predictability and decrease enhancement rates.