In vitro assessment of PEEK and titanium implant abutments: Screw loosening and microleakage evaluations under
Jordi Ortega-Martínez1, Luis M Delgado2, Mónica Ortiz-Hernández3
1Assistant Professor, School of Dentistry, Universitat Internacional de Catalunya, Barcelona, Spain.
Statement Of Problem:
Polyetheretherketone (PEEK) has been advocated to replace metal components in dentistry. Although PEEK is a high-performance polymer with a white color, adequate biological response, and resistance to fracture, data to support PEEK as an alternative material for implant abutments are lacking.
Purpose:
The purpose of this in vitro study was to assess the mechanical and functional properties of PEEK implant abutments as a nonmetallic alternative to titanium abutments, which presents esthetic limitations and greater difficulty to customize clinically.
Material And Methods:
Implant abutments manufactured by using PEEK (n=24) or titanium grade 5 (n=24) were attached to MIS Implants type M4 3.75×16 mm with an internal screw tightened to 25 Ncm. Screw loosening and microleakage was assessed by submersion in a 2% methylene blue solution for 48 hours at 37 °C. Both groups were compared with and without applying dynamic loading; fatigue testing was performed following the International Organization for Standardization (ISO) 14801:2016 standard. All specimens were observed under a stereomicroscope at ×8 magnification. Statistically significant differences among the PEEK and titanium implant abutments were tested with 2-factor ANOVA and the chi-square analysis for nonpaired and paired data, respectively (α=.05).
Results:
The implant abutments made of titanium were better in all mechanical tests. The torque loss of titanium abutments was approximately 10%, while PEEK showed a significantly higher (P<.05) torque loss of up to 50%. Moreover, 91.6% of the titanium abutments did not present microleakage, while there was no specimen of PEEK abutments without microleakage, once subjected to dynamic loading (P<.05).
Conclusions:
Titanium implant abutments (Ti6Al4V) were better in all tests performed. However, PEEK abutments may be suitable for long-term interim restorations, especially in the anterior area, in patients without parafunction.