A Novel Rescue Technique for Difficult Intubation and Difficult Ventilation
Published on: January 17, 2011
Does the minimal occlusive volume technique provide adequate endotracheal tube cuff pressure to prevent air leakage?:
Ha Yeon Park1, Mina Kim1, Junyong In1
1Department of Anesthesiology and Pain Medicine, Dongguk University Ilsan Hospital, Goyang, Korea.
Background:
Methods of determining proper endotracheal tube (ETT) cuff pressure to prevent air leakage include the minimal occlusive volume (MOV) technique, which uses auscultation, and the spirometer technique, which directly measures inspiratory and expiratory breathing volumes. Spirometers may measure even small air leakage, therefore, the spirometer technique requires a higher cuff pressure than the MOV technique to completely seal the airway. This study aimed to evaluate the difference in cuff pressure between the two techniques used to seal the airway.
Methods:
Thirty-five female patients were intubated using an ETT with a cuff, and cuff inflation was performed with both techniques at a 10-min interval in random order-the MOV technique and then the spirometer technique or vice versa. The cuff pressure was measured at each period.
Results:
The cuff pressures were 16.7 ± 4.4 cmH2O and 18.7 ± 5.2 cmH2O for the MOV and spirometer techniques, respectively. The cuff pressure for the spirometer technique was 2.0 cmH2O higher than that for the MOV technique and this difference was statistically significant (95% confidence interval, 0.7-3.3; P = 0.003). Considering the upper end (3.3 cmH2O) of the 95% confidence interval and the size of one scale unit (2.0 cmH2O) of a manometer, the difference in cuff pressure was up to 4 cmH2O in practice.
Conclusions:
Even though the air leakage sound disappears on auscultation, unlike the previous recommendation, the airway sealing would be completed only by increasing the cuff pressure by approximately 4 cmH2O.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...