Technical Note: A simple method for measuring the slice sensitivity profile of iteratively reconstructed CT images
Akihiro Narita1, Masaki Ohkubo1, Takahiro Fukaya1,2
1Graduate School of Health Sciences, Niigata University, 2-746 Asahimachi-dori, Chuo-ku Niigata, 951-8518, Japan.
A new non-slanted edge method accurately measures the slice sensitivity profile (SSP) in computed tomography (CT) iterative reconstruction (IR) images. This simpler approach validates the existing slanted edge method for routine use.
Area of Science:
- Medical Imaging Physics
- Radiological Physics
- Image Reconstruction Algorithms
Background:
- The American Association of Physicists in Medicine (AAPM) Task Group 233 (TG233) established a method for measuring the slice sensitivity profile (SSP) in computed tomography (CT) images reconstructed with iterative reconstruction (IR) algorithms.
- The TG233 method involves a slanted phantom plane edge to create an oversampled edge-spread function (ESF).
Purpose of the Study:
- To investigate the validity of a non-slanted edge method for obtaining a fine-sampled ESF directly from CT images.
- To determine if a simpler, non-slanted approach can accurately measure SSPs, particularly for IR algorithms.
Main Methods:
- A non-slanted edge method was proposed, positioning the phantom edge perpendicular to the z-axis and reconstructing images with a 1-mm slice thickness and 0.1-mm increment.
- The SSP was calculated by differentiating the ESF generated from the mean CT values plotted against slice position.
- Methodology was validated by comparing SSPs from the non-slanted edge method, the TG 233-recommended slanted edge method, and a standard coin method using filtered back projection (FBP) and two IR algorithms.
Main Results:
- For FBP images, the non-slanted edge method demonstrated good agreement with both the coin and slanted edge methods.
- For IR algorithms, SSP measurements using the non-slanted edge method showed strong agreement with those obtained via the slanted edge method.
Conclusions:
- The non-slanted edge method for SSP measurement in CT images is valid.
- The proposed method offers simplicity and practicality, enabling routine and accurate SSP determination for both FBP and IR algorithms.