Real-time assessment of potential peak local specific absorption rate value without phase monitoring: Trigonometric
Ettore Flavio Meliadò1,2,3, Alessandro Sbrizzi1,2, Cornelis A T van den Berg2,4
1Department of Radiology, University Medical Center Utrecht, Utrecht, the Netherlands.
Purpose:
Multi-transmit MRI systems are typically equipped with dedicated hardware to sample the reflected/lost power in the transmit channels. After extensive calibration, the amplitude and phase of the signal at the feed of each array element can be accurately determined. However, determining the phase is more difficult and monitoring errors can lead to a hazardous peak local specific absorption rate (pSAR10g ) underestimation. For this purpose, methods were published for online maximum potential pSAR10g estimation without relying on phase monitoring, but these methods produce considerable overestimation. We present a trigonometric maximization method to determine the actual worst-case pSAR10g without any overestimation.
Theory And Method:
The proposed method takes advantage of the sinusoidal relation between the SAR10g in each voxel and the phases of input signals, to return the maximum achievable SAR10g in a few iterations. The method is applied to determine the worst-case pSAR10g for three multi-transmit array configurations at 7T: (1) body array with eight fractionated dipoles; (2) head array with eight fractionated dipoles; (3) head array with eight rectangular loops. The obtained worst-case pSAR10g values are compared with the pSAR10g values determined with a commonly used method and with a more efficient method based on reference-phases.
Results:
For each voxel, the maximum achievable SAR10g is determined in less than 0.1 ms. Compared to the reference-phases-based method, the proposed method reduces the mean overestimation of the actual pSAR10g up to 52%, while never underestimating the true pSAR10g .
Conclusion:
The proposed method can widely improve the performance of parallel transmission MRI systems without phase monitoring.
More Related Videos
10:46A Method of Trigonometric Modelling of Seasonal Variation Demonstrated with Multiple Sclerosis Relapse Data
Published on: December 9, 2015
06:21Diffuse Optical Spectroscopy for the Quantitative Assessment of Acute Ionizing Radiation Induced Skin Toxicity Using a Mouse Model
Published on: May 27, 2016
Related Concept Videos
One-Compartment Open Model: Wagner-Nelson and Loo Riegelman Method for ka Estimation
On...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Absorption of Radiation
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...